Erika Lewis, Lynn Fors, William J. Tharion; Interrater and Intrarater Reliability of Finger Goniometric Measurements. Am J Occup Ther 2010;64(4):555–561. https://doi.org/10.5014/ajot.2010.09028
Download citation file:
© 2021 American Occupational Therapy Association
Intrarater and interrater reliability were determined for middle finger range of motion (ROM) using the Rolyan finger goniometer. Seven raters measured ROM 3 times (trials) on 20 healthy adults. Intraclass reliability analysis and analysis of variance was used to assess the consistency and stability of measures. Level of significance was set at p ≤ .05. Significant differences (p < .05) existed between raters for metacarpophalangeal, proximal interphalangeal, and distal interphalangeal active finger ROM measurements and for metacarpophalangeal and proximal interphalangeal passive range of motion. Intrarater reliability ranged from .43 to .99 (single measure intraclass coefficient). Interrater reliability ranged from .24 to .95, with passive measures lower than active measures. These results provide clinicians with evidence for using this specific goniometer and issues associated with various therapists assessing finger ROM.
MCP joint: fulcrum: dorsally over MCP joint
Long arm: over dorsal midline of proximal phalanx
Short arm: over dorsal midline of metacarpal
PIP joint: fulcrum: dorsally over PIP joint
Short arm: over dorsal midline of middle phalanx
DIP joint: fulcrum: dorsally over DIP joint
Long arm: over dorsal midline of middle phalanx
Short arm: over dorsal midline of distal phalanx.
Grahic Jump Location